FEATURES
1-50 MHz
Unity gain
Low VSWR
High isolation
High output: +18 dBm
Low phase noise: -147 dBc/Hz @ 1 Hz
-171 dBc/Hz @ 10 kHz
Low distortion: -45 dBc
Low temperature coefficient: 1.5 ps/ºC
APPLICATIONS
Atomic frequency standards
Atomic time scales
High performance testing facilities
Laboratory frequency distribution
Reference frequency distribution
All tests done at 5 MHz and +13 dBm input unless otherwise specified
SPECIFICATIONS
PARAMETER CONDITIONS MIN TYP MAX UNITS
Maximum Input level 1 dB compression 17 18 - dBm
Minimum Input level No fault 7 8 - dBm
Bandwidth +/- 1 dB 1 - 50 0.5- 65 - MHz
Gain @ 5 MHz - 0 +/- 0.5 dB
Impedance Output - 50 - Ohms
Return loss
Input (S11) 5 MHz
Output (S22) 5 MHz
-
-
-25
-35
-20
-30
dB
Distortion +13 dBm - -45 -42 dBc
Isolation Output to output
Output to input
130
140
140
145
-
-
dB
Isolation Input 1 to input 2 - 110 - dB
Switching time Autoswitch mode - 0.5 1 µs
Phase noise
Referred to the Input
1 Hz
10 Hz
1 kHz
10 kHz
-
-
-
-
-147
-157
-167
-171
-145
-155
-166
-170
dBc/Hz
Temperature-delay
coefficient
0 - 50 ºC - 1.5 3 ps/ºC